Laboratorium instrumenten, Petroleum testapparatuur, Temperatuur, Kalibratie apparatuur, Kalibratie Standaarden, Reactoren & automatisatie, Stockverkoop toestellen, Spare parts At the top of technological performance,
it’s the service that makes all the difference.

labo instruments optimus instruments

As market leader in Petroleum Analyzers, Optimus Instruments provides innovative and cost-effective techniques for physical property testing, distillation and element analysis for almost 30 years. Great instruments combined with excellent after sales service and our worldwide unique ISO17025 accredation for verification makes Optimus Instruments a partner you can count on.

0115-013-001 - OPTICPP
POUR POINT ANALYZER (ASTM D97, D5950)

0115-013-001 - Pour Point Analyzer (ASTM D97, D5950)

Combining a revolutionary patented built-in cooling system and highly precisedetection mechanism, the ISL OptiCPP performs pour and/cloud point testing of any petroleum product, down to -95C (-139F), in accordance with all the international standards. Its unique cooling technology eliminates external liquid connections, heat and noise but also the maintenance associated with external cooling.100% configurable, OptiCPP accommodates your unique testing requirements, yet also provides strict compliance to international standard test methods. Automation enables unattended operation with excellent reproducibility and andrepeatability.The OptiCPP is ready to work whenever and wherever you are; simply plug in the electrical cord, and begin testing. The unit can be powered from car battery with an optional adapter. Pour point is detected by tilting the sample no turning, pressing or twisting as an optical surface. The detection system precisely monitors the movement of the specimen surface whereas optics detect cloudpoint in the specimen when wax crystals first appear.

FEATURES & BENEFITS
Precise and reliable analysis performance
Ergonomic cordless cloud point and pour point heads
Powerful and flexible operation with ultra-low temperature capabilities
Enhanced calibration features for full compliance and quality assurance
Multiple options to calibrate temperature probes


Technical details:


In accordance with:
ASTM D5950

In correlation to:
ASTM D97
ISO 3016
IP 15
JIS K2269

Detection:
Cloud Point: optical

Cooling:
Internal cooling system; stepped cooling according to method or user defined; up to 20 steps can be customized

Test Interval:
Pour Point: programmable 1 to 5°C in steps of 0.5°C

Temperature Range:
Jacket: -105 to +55°C (-157 to +131°F)
Sample: -95 to +51°C * (-139 to 123°F)
samples having EPP higher than +35°C have to be preheated externally

Temperature Measurement:
Jacket: 0.5°C accuracy; Pt100 metal probe sample:
0.1°C accuracy; Pt100 glass (standard) or metal (optional) probe engraved with serial number

Calibration:
Automatic temperature calibration routine 20-point temperature probe correction table Programmable calibration frequency

Utility Requirements:
Electrical: 100–240 VAC, 150W, 50/60 Hz
Dimensions: 30 cm x 65 cm x 34 cm (W x D x H)
Weight: 28 kg

Compatible with IRIS software