As market leader in Petroleum Analyzers, Optimus Instruments provides innovative and cost-effective techniques for physical property testing, distillation and element analysis for almost 30 years. Great instruments combined with excellent after sales service and our worldwide unique ISO17025 accredation for verification makes Optimus Instruments a partner you can count on.

0119-012-001 - OPTIMPP
AUTOMATIC MINI CLOUD AND POUR POINT ANALYZER (ASTM D7346)

Delivery To be confirmed.

0119-012-001 - Automatic mini Cloud and Pour point analyzer (ASTM D7346)

Designed to provide highly accurate cloud and pour point temperatures of petroleum products during cooling, ISL’s automated Mini Cloud Pour Point Analyzer is compact, portable and delivers fast results (30C in less than 30 minutes, and quick pour point correlation with no flow @ 0.1C determination).
OptiMPP’s powerful built-in cooling system with selectable cooling profiles requires no special programming or prior cloud and pour point measuring test knowledge.


BENEFITS:

Designed to provide highly accurate cloud and pour point temperatures of petroleum products during cooling, ISL’s automated Mini Cloud Pour Point Analyzer is compact, portable and delivers fast results (30C in less than 30 minutes, and quick pour point correlation with no flow @ 0.1C determination).
OptiMPP’s powerful built-in cooling system with selectable cooling profiles requires no special programming or prior cloud and pour point measuring test knowledge.


Technical details:


METHODS:

Test Methods : ASTM D7346 : No flow point
Correlaton with Pour point: ASTM D97, D5950, IP 15, ISO 3016,
Correlation with Cloud point:: ASTM D2500, D5771, D5772, IP 219, IP 444, ISO 3015
NEW : ASTM D7689-11 Test Method for cloud point, Mini method.


TECHNICAL SPECIFICATIONS:

Cloud Point Detection: Optical detection
Pour Point Detection: Pressure diferential
Cooling Detection: Integrated cooling system
Test Intervals: Continuous 0.1°C resolution

Temperature Range
Sample temperature +45°C to -95°C (to -139°F)
Sample preheating max. 60°C (140°F)
Temperature Measurement °C or °F

IRIS Software:Run Control, Results, Calibration, Reports, Quality Control
Password Security: Multi-level password protection

Calibration
Automatic temperature calibration routine
5-point temperature probe correction table
Programmable calibration frequency

Documentation
On-screen and printed reports (USB/RS printer) Connections to PC or RS (delivered as standard)
Up to 200 results in memory; unlimited storage with PC
Enhanced data management with PAC IRIS Software
Standard Methods: D7346-15, D7689-17

Electrical Connection: 90–240 VAC, 300W, 50/60 Hz
Weight: 30 kg (66 lbs)
Dimensions W x D x H: 25.4 x 60 x 27.4 cm (1” x 23.6 x 10.8 inches)